Table 1.
Different Types of CE318 Instruments Used in the Laboratory Intercalibration.
Inst. | Type | Band Center Wavelength (nm) | Gain-Corrected Viewing Solid Angle (sr) | FOV(°) | Gain Type |
---|
No. 542 | NE | 340, 380, (440), (500), (675), (870), 936, (1020); 1020i, (1640) | 0.05081, 0.04930(10 and 15 December 2010; 16 and 25 May 2012) | 1.27, 1.25 | new |
No. 50 | STD | 340, 380, (440), (675), (870), 936, (1020) | 0.28184(18 April and 02 May 2011;03 May and 29 March 2012) | 2.23 | old |
No. 116 | OP | (440), (675), (870), 936, (1020) | 0.27771(05 and 12 March 2010; 16 and 25 May 2012) | 2.28 | old |
No. 356 | NP | (440), (675), (870), 936, (1020) | 0.05113(22 and 28 September 2011; 16 and 12 June 2012) | 1.28 | new |
No. 719 | NE | 340, 380, (440), (500), (675), (870), 936, (1020); 1020i, (1640) | 0.04853, 0.04970(16 May and 13 June 2012) | 1.26, 1.27 | new |
Note: the “type” column shows internal names for CE318 types used in the PHOTONS group. The “band” column shows the nominal center wavelength, while numbers with parentheses mean that those bands are allowed to perform sky (aureole and dark-sky) measurements. The 1020i is a second 1020 nm band using independent optical path together with a 1640 nm band, and thus are characterized by two viewing solid angles and FOVs. The date in the column of viewing solid angles denotes calibration times of
and
pairs used in the
calculation [
12]. In the case of No. 719, only one calibration pair is available because it has had only one calibration up until now. In the last column, the different gain type determines which kind of gain converting formula will be used in Eq. (
11).
Table 2.
Parameters of the NE Type CE318 No. 542 (the Reference Instrument)
Bands (nm) | () 10 May 2009 | (sr) | (DN) 26 June 2012 | (DN) 16 June 2012 |
---|
1020 | 70.758 | 0.05081 | 13,880 | 17,533 |
870 | 95.953 | 0.05081 | 17,867 | 29,533 |
675 | 150.895 | 0.05081 | 9634 | 34,882 |
440 | 184.316 | 0.05081 | 1569 | 9508 |
500 | 192.932 | 0.05081 | 2039 | 26,514 |
1640 | 23.216 | 0.04930 | 24,418 | 21,075 |
Note: the
values are obtained from the last filter transmittance profile records. The
are measured during the laboratory intercalibration experiment in the unit of digital number (DN). The
values are obtained from field intercalibration by comparison with the AERONET Langley calibrated master instrument No. 44 on 16 June 2012, at Carpentras, France.
Table 3.
and Correction Factors Calculated at 440 nm Bands
Inst. | ASTM2000 | Gueymard | Dif. (%) | Sphere | Blackbody | Dif. (%) |
---|
No. 50 | 1.000 | 1.000 | 0.00 | 1.002 | 1.002 | 0.00 |
No. 116 | 1.008 | 1.008 | 0.00 | 1.011 | 1.010 | 0.10 |
No. 356 | 1.025 | 1.028 | 0.29 | 1.030 | 1.029 | 0.10 |
No. 719 | 1.030 | 1.033 | 0.29 | 1.031 | 1.030 | 0.10 |
Table 4.
Transferring from Master (No. 542) to Secondary Instruments (No. 50, No. 116, No. 356, and No. 719)
Photometer Band (nm) | () | | (sr) | (DN) | (DN) this Work | (DN) AERONET | Dif. (%) |
---|
No. 50 (STD) | 15 April 2011 | | | 26 June 2012 | 26 June 2012 | 17 June 2012 | |
1020 | 70.595 | 1.0000 | 0.28184 | 13,755 | 21,125 | 20,916 | 1.0 |
870 | 95.945 | 0.9970 | 0.28184 | 10,483 | 26,264 | 26,795 | |
675 | 151.277 | 1.0010 | 0.28184 | 5531 | 23,359 | 23,437 | |
440 | 184.242 | 1.0017 | 0.28184 | 766 | 16,742 | 17,022 | |
No. 116 (OP) | 01 January 2000 | | | 26 June 2012 | 26 June 2012 | 02 June 2012 | |
1020 | 70.902 | 1.0002 | 0.27771 | 18,001 | 18,846 | 18,773 | 0.4 |
870 | 95.433 | 0.9920 | 0.27771 | 16,147 | 21,939 | 22,512 | |
675 | 151.276 | 1.0055 | 0.27771 | 9828 | 24,453 | 24,829 | |
440 | 182.855 | 1.0029 | 0.27771 | 1192 | 13,610 | 13,810 | |
No. 356 (NP) | 31 January 2008 | | | 26 June 2012 | 26 June 2012 | 26 August 2012 | |
1020 | 70.927 | 1.0003 | 0.05112 | 15,932 | 20,009 | 19,584 | 2.2 |
870 | 95.603 | 0.9942 | 0.05112 | 20,020 | 32,700 | 32,761 | |
675 | 151.515 | 1.0087 | 0.05112 | 9905 | 17,978 | 18,045 | |
440 | 179.923 | 1.0057 | 0.05112 | 1767 | 22,488 | 22,510 | |
No. 719 (NE) | 16 March 2012 | | | 26 June 2012 | 26 June 2012 | 16 May 2012 | |
1020 | 70.593 | 0.9992 | 0.04852 | 11,701 | 15,466 | 15,228 | 1.6 |
870 | 95.452 | 0.9847 | 0.04852 | 18,288 | 31,170 | 31,772 | |
675 | 151.072 | 1.0028 | 0.04852 | 8982 | 17,076 | 16,815 | 1.6 |
440 | 179.097 | 1.0020 | 0.04852 | 1576 | 10,016 | 10,209 | |
500 | 192.550 | 0.9933 | 0.04852 | 2005 | 27,120 | 26,543 | 2.2 |
1640 | 23.258 | 1.0010 | 0.04970 | 18,813 | 16,122 | 16,369 | |
Table 5.
Error Estimation on the Laboratory Intercalibration Experiment and the Expectation after Improvements
Uncertainties on | This Experiment (%) | Expectation (%) |
---|
| 0.5 | 0.5 |
| | |
| | |
(master) | 1.5 | 0.5 |
Others | 0.5 | 0.3 |
Total on | 2.7 | 1.6 |